Switch off the indicator.
To activate the parameter menu, press the on/off-key for 23 seconds.
The indicator shows: “P 01“.
Press the key.
The indicator shows the current value of parameter P 01.
Use keys6 and5 to change the current value.
Confirm with .
The indicator shows: “P 02“. Parameter P 02 can now be changed in the same way as P 01.
With the 5 key it is possible to scroll through the parameters, until the desired parameter is reached.
When the on/off-key is pressed shortly while a parameter number is shown in the display, the parameter indication will automatically jump back to P 01.
When all desired parameters have been changed, press the key for 3 seconds to leave the parameter menu and to return to the weighing mode.
Listed below are all available parameters. Default parameters are marked with an asterisk (*).
P 01: Not applicable for pallet truck scales. Should be set to 0*.
P 02: Sets smallest multirange graduation.
0.1 = 0.1 kg
0.2 = 0.2 kg
0.5 = 0.5 kg
1 = 1 kg*
2 = 2 kg
5 = 5 kg
10 = 10 kg
20 = 20 kg
50 = 50 kg
P 03: Sets largest multirange graduation.
0.1 = 0.1 kg
0.2 = 0.2 kg
0.5 = 0.5 kg
1 = 1 kg*
2 = 2 kg
5 = 5 kg
10 = 10 kg
20 = 20 kg
50 = 50 kg
Attention: P 02 must always have a lower value than P 03. If, by mistake, a higher value is set, then the indicator automatically sets the same value for both parameters, and the multi-interval function is not used.
P 04: Sets the number of graduations per multirange, per 100 graduations. Available values 0000 – 9900.
Attention: when P 02 and P 03 have equal values, and the multirange is therefore inactive, the value for P 04 is represented as ‘’----“*.
P 05: Sets the weighing capacity (full scale), per 100 kg. Available values 00000 – 99900. Default is 2000*.
P 06: Sets the motion detection, in graduations per second. When the internal value varies per second more than the number of set graduations, the load stability pointer will not light up in the display.
0 = 0.5 graduation
1 = 1 graduation*
2 = 2 graduations
3 = 4 graduations
P 07: Not defined.
P 08: Sets auto-shut off time in minutes. Available values 1- 99 ( 0 = inactive). Default is 3*.
P 09: Sets number of wires per load cell cable.
4 = 4 wires*
P 10: Turns off automatic zero correction.
0 = off
1 = on*
P 11: Not defined.
P 12: Start-up units.
0 = kg
1 = lb
2 = kg/lb*
3 = lb/kg
The start-up unit is also the calibration unit.
P 13 - 16: Not defined.
P 17: Not applicable for pallet truck scales. Should be set to 0*.
P 18 - 19: Undefined.
P 20: Sets baud rate. Available values: 600/1200/2400/4800/9600*.
P 21: Sets the number of data bits. Available values: 7/8*.
P 22: Sets parity.
E = even
- = none*
0 = odd
P 23: Sets number of stop bits. Available values: 1*/2.
P 24: Undefined.
P 25: Sets RS-board function.
0 = standard
1 = standard with printer
2-7 = not in use
P 26: Sets number of line feeds after total print. Available values: 0-7. Default is 5*.
P 27 - 89: Undefined.
P 90: Reset to factory settings. When after selection of parameter P 90 the ¿ key is pressed, all parameter settings are automatically reset to the factory settings.
P 91: Undefined.
P 92: Sets the low battery auto shut off function.
0 = automatic shut off after low battery indication disabled
1 = shut off after 1 minute enabled
P 93: Not applicable. Should be set to 1*.
P 94 - 98: Undefined.
P 99: Shows software version of the indicator.